Articles |
OBJECTIVE: To examine the effect of abnormal cardiac index on the accuracy of measurement of oxygen saturation by pulse oximetry. METHODS: Forty-six patients (mean age, 49 years) in a 9-bed medical ICU were studied. Measurements of oxygen saturation obtained with pulse oximeters and with a functional cooximeter were collected at baseline and 4, 8, 16, 24, 32, 40, and 48 hours later. Hemodynamic and cardiopulmonary parameters were recorded. RESULTS: The Bland-Altman technique yielded upper and lower limits of agreement of 2.53% and -7.11%. Most (95.7%) of the differences between the measurements of oxygen saturation obtained with the 2 methods were within these limits, although some of these differences may be clinically unacceptable. The bias was -2.29%, and the precision was 2.41%. The clinical conditions associated with inaccurate tracking of saturation by pulse oximetry across the range of actual arterial oxygen saturation values were abnormal cardiac index, partial pressure of carbon dioxide, heart rate, and pulmonary capillary wedge pressure. CONCLUSIONS: In patients with abnormal cardiac index, the pulse oximeter measurements exceeded the actual oxygen saturation by up to 7%. Pending prospective studies, clinicians should be aware that when certain cardiopulmonary parameters are abnormal, the margin of error in measurements of oxygen saturation obtained with a pulse oximeter may be greater than when those parameters are normal.
This article has been cited by other articles:
![]() |
T. W. Rice, A. P. Wheeler, G. R. Bernard, D. L. Hayden, D. A. Schoenfeld, L. B. Ware, and for the National Institutes of Health, National He Comparison of the SpO2/FIO2 Ratio and the PaO2/FIO2 Ratio in Patients With Acute Lung Injury or ARDS Chest, August 1, 2007; 132(2): 410 - 417. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Barker, T. Lang, H. Hager, B. Steinlechner, K. Hoerauf, M. Zimpfer, and A. Kober The Influence of Stellate Ganglion Transcutaneous Electrical Nerve Stimulation on Signal Quality of Pulse Oximetry in Prehospital Trauma Care Anesth. Analg., May 1, 2007; 104(5): 1150 - 1153. [Abstract] [Full Text] [PDF] |
||||
![]() |
S.-Y. P. K. Shiao and C.-N. Ou Validation of Oxygen Saturation Monitoring in Neonates Am. J. Crit. Care., March 1, 2007; 16(2): 168 - 178. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. S. Balan, G. Fiskum, J. Hazelton, C. Cotto-Cumba, and R. E. Rosenthal Oximetry-Guided Reoxygenation Improves Neurological Outcome After Experimental Cardiac Arrest Stroke, December 1, 2006; 37(12): 3008 - 3013. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |