American Journal of Critical Care. 2007;16: 336-346

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Respond to This Article
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Related articles in AJCC
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Right arrow Take the CE Test
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lacara, T.
Right arrow Articles by Prasnikar, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lacara, T.
Right arrow Articles by Prasnikar, M.

CE Article

Comparison of Point-of-Care and Laboratory Glucose Analysis in Critically Ill Patients

By Teresita Lacara, RN, BSN, Caroline Domagtoy, RN, BSN, Donna Lickliter, RN, Kathy Quattrocchi, RN, BSN, Lydia Snipes, RN, Joánne Kuszaj, RN, MSN, CCRN and MaryClare Prasnikar, RN, MSN, CCRN. Teresita Lacara, Caroline Domagtoy, Donna Lickliter, and Lydia Snipes are staff nurses; Kathy Quattrocchi is a team leader; Joánne Kuszaj is a clinical manager; and MaryClare Prasnikar is a cardiovascular clinical nurse specialist in the medical-surgical intensive care unit at Rex Healthcare in Raleigh, NC.

Corresponding author: Teresita Lacara, RN, BSN, Intensive Care Unit, Rex Healthcare, 4424 Lake Boone Trail, Raleigh, NC 27607 (e-mail: teresita.lacara{at}rexhealth.com).


    Abstract
 Top
 Abstract
 Materials and Methods
 Results
 Discussion
 Clinical Implications
 Conclusions
 References
 
Background Blood for point-of-care analysis of glucose levels is often obtained from different sources (fingerstick, arterial or central venous catheter).

Objectives To examine agreement between point-of-care and laboratory glucose values and to determine effects of hematocrit, serum carbon dioxide, and mean arterial pressure on the accuracy of point-of-care values.

Methods Point-of-care values were compared with laboratory values. In 49 critically ill patients, blood was obtained first from a catheter for laboratory testing and then from the catheter and via fingerstick for point-of-care testing. Bias, precision, and root-mean-square differences were calculated to quantify differences in values between the 2 methods. A t test was used to determine differences in values between each point-of-care blood source and the laboratory value. Multiple regression analysis was used to determine if serum level of carbon dioxide, hematocrit, and/or mean arterial pressure significantly contributed to the difference in bias and precision for the point-of-care blood sources.

Results Mean laboratory glucose level was 135 (SEM 5.3, range 58–265) mg/dL. In point-of-care testing, bias ± precision and root-mean-square differences were 2.1 ± 12.3 and 12.35, respectively, for fingerstick blood and 0.6 ± 10.6 and 10.46 for catheter blood. Values for point-of-care and laboratory tests did not differ significantly. For catheter samples, hematocrit and serum carbon dioxide contributed significantly to difference scores between point-of-care and laboratory values (P < .001).

Conclusions Glucose values for point-of-care samples did not differ significantly from laboratory values. For catheter samples, hematocrit and serum carbon dioxide levels accounted for the difference between point-of-care and laboratory glucose values.

Notice to CE enrollees:A closed-book, multiple-choice examination following this article tests your understanding of the following objectives:
  1. Discuss use of point-of-care laboratory testing for glucose analysis in critically ill patients.
  2. Identify factors that can influence the accuracy of point-of-care glucose analysis during critical illness.
  3. Describe how hematocrit values can influence glucose analysis in critically ill patients.

To read this article and take the CE test online, visit www.ajcconline.org and click "CE Articles in This Issue." No CE test fee for AACN members.


Abnormal blood glucose levels are common in critically ill patients and increase the risk for complications such as infection, metabolic problems, and/or cerebral damage.15 Frequent monitoring of blood glucose levels and aggressive management of hyperglycemia can decrease these complications and mortality.3,613 Although laboratory analysis is the most accurate method for evaluating glucose levels, because of cost and time delays, bedside point-of-care (POC) testing is often used to determine glucose levels when frequent monitoring of glucose is important. Although POC glucose meters were designed to be used with capillary blood obtained from a fingerstick, clinicians often obtain blood for POC testing from arterial or central venous pressure (CVP) catheters.

Several investigators1423 have evaluated the accuracy of fingerstick POC glucose testing compared with that of laboratory glucose analysis, but clinical studies2427 on the accuracy of using blood from arterial or CVP catheters for POC glucose determination are limited. None of these studies2427 included evaluations of the impact of biochemical derangements, such as altered pH, on the accuracy of glucose meters when blood from arterial or CVP catheters was used, and the impact of abnormal hematocrit values25 or poor tissue perfusion27 was evaluated in only a single study. In earlier studies14,15,23,2835 with capillary blood, these conditions interfered with the accuracy of POC glucose values.

The purpose of this study was to compare POC glucose meter values of both capillary (fingerstick) and arterial or CVP blood samples with laboratory glucose values in critically ill patients. In addition, we examined whether hematocrit, serum level of carbon dioxide, and mean arterial pressure (MAP) affected the bias of the different sources of blood, because other investigators14,15,23,25,28,29,3134 either found or hypothesized that these physiological variables affected or would affect the accuracy of measurements obtained with glucose meters.


    Materials and Methods
 Top
 Abstract
 Materials and Methods
 Results
 Discussion
 Clinical Implications
 Conclusions
 References
 
This study was conducted in a 394-bed community-based hospital in the southeastern region of the United States. Approval was obtained from the institution’s investigational review board before any data were collected.

Study Design
A method-comparison study design was used to compare glucose values obtained with a POC device and a clinical laboratory analysis method. The dependent variables were the differences between glucose values obtained with the POC testing device (fingerstick and arterial or CVP catheter specimens) and the values obtained with the clinical laboratory method.


Although POC glucose meters were designed to be used with capillary blood obtained from a fingerstick, clinicians often obtain blood for POC testing from arterial or central venous catheters.

 

Sample
A convenience sample of critically ill patients had POC testing done once at the same time a blood sample was collected for glucose determination via the laboratory method. Inclusion criteria included the presence of a CVP or an arterial catheter. Sample size was determined a priori by power analysis (power = 80%, {alpha} = .05, effect size = 0.73 for t test and 0.25 for multiple regression analysis).36 Determinations of effect size were based on the national standard for minimal acceptable accuracy for glucose POC devices of 20% or less variation from laboratory values for glucose concentrations of 75 mg/dL or greater37 (to convert glucose values to millimoles per liter, multiply by 0.05551).

Procedure
In a standard procedure, blood was obtained from the CVP or arterial catheter and placed in a separator vacuum test tube for laboratory glucose testing. The minimum amount of catheter blood discarded before laboratory glucose sampling was 5 mL, a volume that is at least 5 times the catheter dead space. Blood for laboratory testing was analyzed by using standard procedures. Laboratory glucose was determined by using the adapted hexokinase-glucose-6-phosphate method (Dimension Clinical Chemistry System, model RxL, Dade Behring Inc, Deerfield, Illinois). Manufacturer’s specifications indicate less than 5% mean for the coefficient of variation for this model of glucose analyzer.38 Mean time from blood sampling to laboratory analysis is 45 minutes in our hospital.


No significant difference between laboratory and point-of-care glucose values was found.

 

After the CVP or arterial catheter blood was obtained for laboratory testing of glucose, an additional 0.1 mL of blood was obtained from the catheter for glucose measurement with the POC glucose oxidase photometry device (SureStep Pro Hospital Meter and SureStep Pro Hospital Products Test Strips, Johnson & Johnson, Milpitas, California).38 As specified in the manufacturer’s directions, glucose testing was performed by placing a drop of blood on the reagent strip and then placing the strip into the glucose meter for reading. Capillary blood was then immediately obtained by lancing a fingertip to produce a drop of blood, which was then tested for glucose with the POC testing device (finger-stick POC testing) via the same procedure as that used for catheter blood. All POC testing was done immediately after blood was obtained from the catheter or via fingerstick. The same investigator (C.D.) obtained and analyzed all the POC samples.


Hematocrit, serum carbon dioxide level, and blood pressure may affect point-of-care glucose measurement accuracy.

 

Hematocrit, serum carbon dioxide, and MAP values at the time of therapeutic glucose determination were hand logged onto a data sheet for each patient. Hematocrit and serum carbon dioxide levels were analyzed by the hospital clinical laboratory with a Beckman Coulter LH 750 hematology analyzer (Beckman Coulter Inc, Brea, California) and a Dimension RxL Max w/HM and RMS chemistry analyzer (Dade Behring Inc), respectively, according to the manufacturers’ guidelines. MAP was measured directly via a radial artery catheter connected to a pressure transducer (MX9501T TranStar Patient Mount Monitoring Kit, Medex, Dublin, Ohio) and a bedside pressure monitor (Solar 8000, model 415982-005, GE Marquette Medical Systems, Milwaukee, Wisconsin).

Quality control of the glucose meters was done daily according to manufacturer’s directions and included testing both high- and low-quality control reagents.38 Care was taken to ensure that POC test strips were from the same lot number and had not expired. Before data collection, the investigator (C.D.) who would be performing the POC sampling and analysis was trained in the proper use of the POC glucose meter.38

Data Analysis
Data were summarized by using descriptive statistics. Difference scores between both POC glucose values (fingerstick and arterial or CVP) and laboratory glucose values also were calculated for each patient. Mean difference scores, or device bias, and limits of agreement between the POC test values (fingerstick and arterial or CVP) and the laboratory glucose values were calculated by using the Bland-Altman method.3941 Also, t tests were used to determine if differences between the laboratory glucose value and each of the POC (fingerstick and arterial or CVP) glucose values were significant. Multiple regression analysis was used to determine if hematocrit, serum carbon dioxide, and/or MAP accounted for the difference scores between the laboratory value and the POC glucose values. The level of significance for all statistical tests was set at P < .05, with a Bonferoni correction for the multiple t tests.


    Results
 Top
 Abstract
 Materials and Methods
 Results
 Discussion
 Clinical Implications
 Conclusions
 References
 
A total of 49 patients were evaluated. Demographic data for the sample are presented in Table 1Go. Mean age was 66.8 (SEM 2.2) years. Of the 49 patients, 13 had diabetes and 3 had steroid-induced hyperglycemia. Blood for laboratory testing was obtained from an arterial catheter in 42 patients and from a CVP catheter in 7. The ranges and mean values of hematocrit, serum carbon dioxide, and MAP are summarized in Table 2Go.


View this table:
[in this window]
[in a new window]

 
Table 1 Sample characteristics for 49 critically ill patients

 

View this table:
[in this window]
[in a new window]

 
Table 2 Values for selected physiological variables for 49 critically ill patients

 
Laboratory glucose values ranged from 58 to 265 mg/dL; fingerstick and catheter POC glucose values ranged from 52 to 281 mg/dL and from 61 to 263 mg/dL, respectively (Table 3Go). Glucose values were normally distributed. Bias (difference) and precision (limits of agreement) were 2.1 and 12.3, with root-mean-square differences (RMSDs) of 12.35 for the fingerstick POC and laboratory glucose values and 0.6 and 10.6, with RMSD of 10.46 for the catheter POC and laboratory glucose values (Table 3Go, Figure 1Go). Results of the t tests indicated no significant differences between the laboratory glucose value and the POC glucose values (fingerstick POC: t48 = 1.21, P = .23; catheter POC: t48 = –0.40, P = .69).


View this table:
[in this window]
[in a new window]

 
Table 3 Glucose values, bias, precision, and root-mean-square of the difference scores for 49 critically ill patientsa

 

Figure 1
View larger version (34K):
[in this window]
[in a new window]

 
Figure 1 Bland-Altman graphs depicting the differences between the laboratory glucose value and the point-of-care (POC) glucose values determined with blood from a fingerstick (A) or an arterial or a central venous pressure (CVP) catheter (B) for 49 patients.

To convert glucose values to millimoles per liter, multiply by 0.05551.

 
Multiple regression analysis indicated that hematocrit and serum carbon dioxide levels were significant contributors to difference scores between the laboratory and the catheter POC methods (F3,45 = 8.17, P < .001; Table 4Go). MAP did not significantly account for the difference scores. No significant contributors were found for difference scores between the laboratory and fingerstick POC analysis methods (F3,45 = 2.56, P = .07).


View this table:
[in this window]
[in a new window]

 
Table 4 Multiple regression analysis of POC and laboratory glucose difference scores for 3 independent variables

 
Because the number of samples obtained via the CVP catheters was so small, the 7 POC blood samples obtained via a CVP catheter were removed from the study. The bias and precision (differences and limits of agreement; Table 3Go, Figure 2Go) and results of multiple regression analysis (Table 4Go) for the arterial POC and laboratory glucose values (n = 42) were similar to those for the entire sample (N = 49), with the exception of serum carbon dioxide. Multiple regression analysis for the fingerstick POC method indicated that serum carbon dioxide level was a significant contributor to the difference scores between the laboratory and arterial POC analysis methods (P = .04; Figure 3Go).


Figure 2
View larger version (33K):
[in this window]
[in a new window]

 
Figure 2 Bland-Altman graphs depicting the differences between laboratory glucose value and the point-of-care (POC) glucose value obtained with blood from a fingerstick (A) or an arterial catheter (B) for 42 patients.

To convert glucose values to millimoles per liter, multiply by 0.05551.

 

Figure 3
View larger version (32K):
[in this window]
[in a new window]

 
Figure 3 Scattergrams of point-of care (POC) and laboratory glucose differences for significant multiple regression variables, hematocrit and serum carbon dioxide level, in 42 patients with POC testing of arterial blood. A, hematocrit and arterial POC; B, serum carbon dioxide and arterial POC; C, serum carbon dioxide and fingerstick POC.

To convert glucose values to millimoles per liter, multiply by 0.05551.

 

    Discussion
 Top
 Abstract
 Materials and Methods
 Results
 Discussion
 Clinical Implications
 Conclusions
 References
 
Blood obtained from arterial and CVP catheters had slightly smaller bias and precision values than did blood from a fingerstick sample. The differences between laboratory and POC glucose values, however, were not statistically significant. In addition, hematocrit and serum carbon dioxide levels significantly accounted for the differences between the laboratory and POC testing values.

Although obtaining blood for POC glucose testing from arterial or venous catheters instead of using the fingerstick method is common in critical care units, this practice has been examined in only a few published studies.2427 Only 2 of these studies24,25 had sufficient detail and clarity for adequate interpretation of the results. Small sample size,24 evaluation of devices no longer used in clinical practice,26 and inappropriate statistical analysis25,26 limit the generalizability of the studies to clinical practice.

In a study of 50 postoperative cardiac surgical patients, Maser et al25 found a significant difference between the mean glucose values obtained with an arterial POC device (Accu-Check II, Boehringer-Mannheim) and the arterial sample analyzed by using the laboratory method: 249 (SD 12) mg/dL and 219 (SD 12) mg/dL, respectively. Mean POC values with fingerstick samples, 210 (SD 12) mg/dL, were lower than the laboratory mean value. Maser et al hypothesized that the large differences between the various methods for glucose analysis might have been related to low body temperature and the effect on the enzymatic reactions on the test strips rather than to the source of the blood for analysis or to systematic error.

In the study by Maser et al,25 all POC testing was performed by the same investigator to minimize user error. A major limitation of the study, which makes interpretation of the data difficult, is that difference scores between the POC value and the laboratory reference value were not computed and graphed according to the Bland-Altman method.39,40 Experts in technology assessment consider the Bland-Altman method the best way to examine the level of agreement between 2 medical devices.3941 Correlational methods are inappropriate for analysis in these types of studies because, with those methods, the degree of relationship rather than the difference between 2 variables is examined.

Unlike Maser et al,25 Ray et al24 found no difference between arterial glucose values obtained with a POC device (One Touch Profile, Lifescan, Johnson & Johnson) and values obtained with the laboratory reference standard. In that study,24 blood was obtained from 10 critically ill patients in a total of 105 sampling periods. The results of our study were similar to those of Ray et al, but we used a different POC device (SureStep Pro, Lifescan, Johnson & Johnson).


When point-of-care measures are used, a low hematocrit level may result in overestimates of blood glucose.

 

Hematocrit has long been known to affect the accuracy of POC glucose analysis.42,43 Despite manufacturers’ operating instructions for POC devices that suggest limiting use of the devices to clinical situations in which hematocrit levels are within a specific range of values, typically 25% to 55% or 60%, POC devices often are used clinically without regard for hematocrit level. Similar to our finding, the results of several studies14,15,25,28,29,3133 of different glucose meters have indicated that lower than normal hematocrit values (<30% to <35%) result in overestimates of laboratory glucose levels when the POC method is used, whereas hematocrit values higher than normal (>45%) result in underestimates of laboratory values. Although the mechanism for these differences is not known, various hypotheses have been proposed to explain the impact of abnormal hematocrit levels on POC testing: altered viscosity of the blood, prevention of plasma from reaching the reaction surface of the test strip, change in diffusion kinetics, and/or increased packed red cell volume and displacement of plasma volume leading to insufficient plasma volume for accurate testing.15


Any single point-of-care glucose value for titration of insulin should be used cautiously.

 

In earlier studies14,15,25,28,29,3133 on the effect of hematocrit on POC testing, the investigators used fingerstick (capillary) blood and found a negative bias of glucose values with increasing levels of hematocrit. We did not find that hematocrit significantly explained the difference between fingerstick POC glucose values and laboratory values. This difference in findings may be due to differences in device performance, because the POC devices in the earlier studies differed from the POC device we used. Another possible explanation is that hematocrit ranges in our study were more homogeneous than were the ranges in the other studies. Although hematocrit values ranged from 20% to 53% in our study, mean hematocrit was 32% with a small SEM (0.8%). Few of our subjects had hematocrit values outside the manufacturers’ recommended range for optimal device functioning (25%–60%).

Another reason for the different results might be the changes made in POC devices used today to improve the accuracy of measurement of capillary blood glucose levels in patients with abnormal hematocrit levels. This last reason also would explain why hematocrit levels significantly accounted for differences between POC glucose values obtained with CVP or arterial blood and the laboratory value in our study but not for the differences between POC values with capillary (fingerstick) blood and the laboratory value. Any improvements in device performance made to correct the effect of hematocrit on glucose determinations when the POC device is used with capillary samples could then cause an overcorrection if the source of blood used for POC testing is not capillary blood.

In addition to the adverse effect of hematocrit levels on the accuracy of POC glucose values, several other factors can interfere with the accuracy of different models of POC glucose meters44: creatinine levels,45,46 bilirubin levels,30 high arterial oxygen levels,45,46 shocklike states,23,34,35,47 and altered pH.31,34 On the basis of anecdotal experience with discrepancies between POC and laboratory glucose values in individual patients with shock states, we decided to evaluate MAP and serum carbon dioxide level, because arterial pH values are not routinely available in our patients. Serum carbon dioxide levels vary in relationship to alterations in blood pH, so we hypothesized that serum carbon dioxide would be associated with differences between POC and laboratory glucose. In all but one of the multiple regression analyses we performed (fingerstick POC glucose values minus laboratory glucose values), serum carbon dioxide significantly accounted for the POC and laboratory difference scores. This finding is similar to, and consistent with, findings in previous studies31,34 on the effect of pH on the accuracy of POC glucose values, because serum levels of carbon dioxide depend on serum pH.

Unlike the findings in previous studies23,34,35,47 of shocklike states, MAP did not significantly explain the difference scores between POC and laboratory glucose values in our study. Because most of our subjects had MAP greater than 60 mm Hg, and thus did not have shocklike states, our findings are not surprising. Future studies should be done in patients with shock to determine if MAP is a significant contributor to POC accuracy.


    Clinical Implications
 Top
 Abstract
 Materials and Methods
 Results
 Discussion
 Clinical Implications
 Conclusions
 References
 
The lack of differences between glucose values obtained via a laboratory analysis and values obtained with the POC device, whether capillary or catheter blood was used, supports the common practice in critical care units of using catheter rather than finger-stick blood for POC testing. With the advent of aggressive glucose management protocols to decrease infection risks in critically ill patients,3,613 blood samples are needed frequently, and use of catheter blood avoids painful needlesticks to obtain capillary blood.

One limitation of our findings is that all of the POC tests were done by the same investigator. This practice, although ideal for research control, is not typical of the way POC testing is usually done. Significant user error can be introduced with multiple users and can make the difference between POC and laboratory values in actual practice situations more pronounced than the differences in our study. Having multiple care providers do the POC testing in a replication of our study would be prudent to determine the impact of the number of testers on the accuracy of POC glucose values. In the other study25 that showed no difference between POC and laboratory glucose values, a single provider also did all the POC tests.

Because of the range of difference scores and the width of the 95% CIs (the ±2 SD lines) of the Bland-Altman graphs (Figures 1Go and 2Go) in our study, we recommend caution in using any individual POC glucose value as a basis for adjusting insulin doses when tight glucose management protocols are being used. Many of the patients in our study had glucose differences of at least 10 mg/dL between the POC and laboratory methods, which may be a large enough difference from the true glucose value to change management decisions when treatment protocols call for narrow ranges for glucose levels.

As shown in other studies,14,15,25,28,29,3133 abnormal hematocrit levels may influence the accuracy of a POC device. In patients with abnormally low hematocrit levels, periodically validating the POC test result by comparing it with a laboratory glucose level would be a prudent approach to avoid situations in which the POC value is an overestimation of the true glucose value.


    Conclusions
 Top
 Abstract
 Materials and Methods
 Results
 Discussion
 Clinical Implications
 Conclusions
 References
 
Our findings validated the practice in critical care units of obtaining blood for POC testing from arterial or venous catheters rather than from a fingerstick source. The bias and precision of glucose POC testing with a fingerstick sample was slightly higher than the bias and precision with samples obtained from an arterial or a CVP catheter, but the differences were not statistically significant. In addition, for hematocrit and serum carbon dioxide levels below normal ranges, POC values with arterial blood tended to be overestimates of laboratory glucose values. In individual situations, for patients with abnormal hematocrit or carbon dioxide levels, care should be taken to verify the accuracy of POC glucose testing by comparing the POC value with the laboratory value of a sample obtained at the same time as the POC sample. Further study is needed to determine if patients with abnormally low MAP have greater discrepancies between POC and laboratory glucose values.


    ACKNOWLEDGMENTS
 
Special thanks to Marianne Chulay, RN, DNSc, FAAN, for assistance with study design, data analysis, and manuscript preparation.

FINANCIAL DISCLOSURES
None reported.

eLetters
Now that you’ve read the article, create or contribute to an online discussion about this topic using eLetters. Just visit www.ajcconline.org and click "Respond to This Article" in either the full-text or .pdf view of the article.

To purchase electronic or print reprints, contact The InnoVision Group, 101 Columbia, Aliso Viejo, CA 92656. Phone, (800) 809-2273 or (949) 362-2050 (ext 532); fax, (949) 362-2049; e-mail, reprints{at}aacn.org.


    REFERENCES
 Top
 Abstract
 Materials and Methods
 Results
 Discussion
 Clinical Implications
 Conclusions
 References
 

  1. Umpierrez GE, Isaacs SD, Bazargan N, You X, Thaler LM, Kitabchi AE. Hyperglycemia: an independent marker of inhospital mortality in patients with undiagnosed diabetes. J Clin Endocrinol Metab. 2002;87:978–982.[Abstract/Free Full Text]
  2. Van den Berghe G, Wouters P, Weekers F, et al. Intensive insulin therapy in critically ill patients. N Engl J Med. 2001;345:1359–1367.[Abstract/Free Full Text]
  3. Pomposelli JJ, Baxter JK III, Babineau TJ, et al. Early postoperative glucose control predicts nosocomial infection rate in diabetic patients. J Parenter Enteral Nutr. 1998;22:77–81.[Abstract]
  4. Capes SE, Hunt D, Malmberg K, Gerstein HC. Stress hyperglycaemia and increased risk of death after myocardial infarction in patients with and without diabetes: a systematic overview. Lancet. 2000;355:773–778.[Medline]
  5. Pierre EJ, Barrow RE, Hawkins HK, et al. Effects of insulin on wound healing. JTrauma. 1998;44:342–345.
  6. Dagogo-Jack S, Alberti KG. Management of diabetes mellitus in surgical patients. Diabetes Spectr. 2002;15:44–48.[Abstract/Free Full Text]
  7. Dellinger EP. Preventing surgical site infections: the importance of timing and glucose control. Infect Control Hosp Epidemiol. 2001;22:604–606.[Medline]
  8. Jacober SJ, Sowers JR. An update of perioperative management of diabetes. Arch Intern Med. 1999;159:2405–2411.[Abstract/Free Full Text]
  9. Malmberg K, Ryden L, Efendic S, et al. Randomized trial of insulin-glucose infusion followed by subcutaneous insulin treatment in diabetic patients with acute myocardial infarction (DIGAMI study): effects on mortality at 1 year. J Am Coll Cardiol. 1995;26:57–65.[Abstract]
  10. Furnary AP, Gao G, Grunkemeier GL, et al. Continuous insulin infusion reduces mortality in patients with diabetes undergoing coronary artery bypass grafting. JThorac Cardiovasc Surg. 2003;125:1007–1021.[Abstract/Free Full Text]
  11. Van den Berghe G, Wouters PJ, Bouillon R, et al. Outcome benefit of intensive insulin therapy in the critically ill: insulin dose versus glycemic control. Crit Care Med. 2003;31:359–366.[Medline]
  12. Clement S, Braithwaite SS, Magee MF, et al. Management of diabetes and hyperglycemia in hospitals [published corrections in Diabetes Care. 2004;27:836, 1255]. Diabetes Care. 2004;27:553–591.[Free Full Text]
  13. Krinsley JS. Effect of an intensive glucose management protocol on the mortality of critically ill adult patients [published correction in Mayo Clin Proc. 2005;80:1101]. Mayo Clin Proc. 2004;79:992–1000.[Medline]
  14. Puntmann I, Wosniok W, Haeckel R. Comparison of several point-of-care testing (POCT) glucometers with an established laboratory procedure for the diagnosis of type 2 diabetes using the discordance rate: a new statistical approach. Clin Chem Lab Med. 2003;41;809–820.[Medline]
  15. Louie RF, Tang Z, Sutton DV, Lee JH, Kost GJ. Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design. Arch Pathol Lab Med. 2000;124;257–266.[Medline]
  16. Tang Z, Du X, Louie RF, Kost GJ. Effects of pH on glucose measurements with handheld glucose meters and portable glucose analyzer for point-of-care testing. Arch Pathol Lab Med. 2000;124:577–582.[Medline]
  17. Velazquez Medina D, Climent C. Comparison of outpatient point of care glucose testing vs venous glucose in the clinical laboratory. P R Health Sci. 2003;22;385–389.
  18. Poirier JY, Le Prieur N, Campion L, Guilhem I, Allannic H, Maugendre D. Clinical and statistical evaluation of self-monitoring blood glucose meters. Diabetes Care. 1998;21:1919–1924.[Abstract]
  19. Fanghanel G, Sanchez-Reyes L, Morales M, et al. Comparative accuracy of glucose monitors. Arch Med Res. 1998;29:325–329.[Medline]
  20. Bohme P, Floriot M, Sirveaux MA, et al. Evolution of analytical performance in portable glucose meters in the last decade. Diabetes Care. 2003;26:1170–1175.[Abstract/Free Full Text]
  21. Harding K. A comparison of four glucose monitors in a hospital medical surgical setting. Clin Nurse Spec. 1993;7:13–16.[Medline]
  22. Aboezz R, Miller D. Accuracy of portable blood glucose monitors. J Am Pharm Assoc (Wash DC). 2005;45:514–516.
  23. Atkin SH, Dasmahapatra A, Jaker MA, Chorost MI, Reddy S. Fingerstick glucose determination in shock. Ann Intern Med. 1991;114:1020–1024.[Medline]
  24. Ray JG, Hamielec C, Mastracci T. Pilot study of the accuracy of bedside glucometry in the intensive care unit. Crit Care Med. 2001;29:2205–2207.[Medline]
  25. Maser RE, Butler MA, DeCherney GS. Use of arterial blood with bedside glucose reflectance meters in an intensive care unit: are they accurate? Crit Care Med. 1994:22;595–599.[Medline]
  26. Pressly KB, Batteiger TH, Barnett DZ, Woodie ME. Use of arterial blood for glucose measurement by reflectance. Nurs Res. 1990;39:371–373.[Medline]
  27. Kanji S, Buffie J, Hutton B, et al. Reliability of point-of-care testing for glucose measurement in critically ill adults. Crit Care Med. 2005;33:2778–2785.[Medline]
  28. Phillipou G, Seaborn CJ, Hooper J, Phillips PJ. Capillary blood glucose measurements in hospital inpatients using portable glucose meters. Aust N Z J Med. 1993;23:667–671.[Medline]
  29. Smith EA, Kilpatrick ES. Intra-operative blood glucose measurements: the effect of haematocrit on glucose test strips. Anaesthesia. 1994;49:129–132.[Medline]
  30. Jain R, Myers TF, Kahn SE, Zeller WP. How accurate is glucose analysis in the presence of multiple interfering substances in the neonate? J Clin Lab Anal. 1996;10:13–16.[Medline]
  31. Tang Z, Lee JH, Louie RF, Kost GJ. Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing. Arch Pathol Lab Med. 2000;124:1135–1140.[Medline]
  32. Wiener K. The effect of haematocrit on reagent strip tests for glucose. Diabet Med. 1991;8:172–175.[Medline]
  33. Barreau PB, Buttery JE. Effect of hematocrit concentration on blood glucose value determined on Glucometer II. Diabetes Care. 1988;11:116–118.[Abstract]
  34. Kilpatrick ES, Rumley AG, Smith EA. Variations in sample pH and pO2 affect ExacTech meter glucose measurements. Diabet Med. 1994;11:506–509.[Medline]
  35. Sylvain HF, Pokorny ME, English SM, et al. Accuracy of fingerstick glucose values in shock patients. Am J Crit Care. 1995;4:44–48.[Abstract]
  36. Cohen J. Statistical Power Analysis for the Behavioral Sciences. Rev ed. New York, NY: Academic Press; 1977:313.
  37. International Organization for Standardization. In Vitro Diagnostic Test Systems Requirements for Blood-Glucose Monitoring Systems for Self-testing in Managing Diabetes Mellitus. Geneva, Switzerland: International Organization for Standardization; 2003. Publication 15197.
  38. SureStep Pro Hospital Meter Product Guidelines. Milpitas, CA: Johnson & Johnson Inc.
  39. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–310.[Medline]
  40. Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 2000;8:135–160.
  41. Chatburn R. Evaluation of instrument error and method of agreement. AANA J. 1996;64:261–268.[Medline]
  42. Consensus statement on self-monitoring of blood glucose. Diabetes Care. 1987;10:95–99.[Medline]
  43. American Diabetes Association. Clinical practice recommendations 1996. Diabetes Care. 1996;19(suppl 1):S1–S118.[Medline]
  44. Using blood glucose meters: minimizing errors, maximizing accuracy. Health Devices. 2004;33:251–256.[Medline]
  45. Kurahashi K, Maruta H, Usuda Y, Ohtsuka M. Influence of blood sample oxygen tension on blood glucose concentration measured using an enzyme-electrode method. Crit Care Med. 1997;25:231–235.[Medline]
  46. Zaloga GP. Beware of errors in blood glucose measurements [editorial]. Crit Care Med. 1997;25:212.[Medline]
  47. Kulkarni A, Saxena M, Price G, O’Leary MJ, Jacques T, Myburgh JA. Analysis of blood glucose measurements using capillary and arterial blood samples in intensive care patients. Intensive Care Med. 2005;31:142–145.[Medline]

Related articles in AJCC:

Clinical Pearls
Mary Jo Grap
AJCC 2007 16: 335. [Full Text]  



This article has been cited by other articles:


Home page
Am J Crit CareHome page
E. A. Mann, H. F. Pidcoke, J. Salinas, C. E. Wade, J. B. Holcomb, S. E. Wolf, T. Lacara, C. Domagtoy, D. Lickliter, K. Quattrocchi, et al.
Accuracy of Glucometers Should Not Be Assumed
Am. J. Crit. Care., November 1, 2007; 16(6): 531 - 532.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Respond to This Article
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Related articles in AJCC
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Right arrow Take the CE Test
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lacara, T.
Right arrow Articles by Prasnikar, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lacara, T.
Right arrow Articles by Prasnikar, M.


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS